Read the docs
TI-2018-07490 Aliases CVE-2018-10601GHSA-9f6x-r29g-cv4j

Stack-based buffer overflow in ICS-CERT IntelliVue Patient Monitors, Avalon…

Priority LOW 36CWE-121Published · Updated

Key facts

Attack vector
Adjacent network
Privileges required
None
User interaction
None
Attack complexity
High
Impact type
—
Confidentiality · integrity · availability
High · Low · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · NVD. Fix status from CNA.

Summary

IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

Why it matters now

EPSS 0.00372 31st percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
ICS-CERTIntelliVue Patient Monitors, Avalon Fetal/Maternal MonitorsThe following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3—No fix listedCNA
  • IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors VendorICS-CERT Affected The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 Fixed in — No fix listed CNA

CPE configurations NVD · 36

CPEVersionsRoleVendor
cpe:2.3:o:philips:intellivue_mp2_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mp2:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_x2_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_x2:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mp30_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mp30:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mp50_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mp50:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mp70_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mp70:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
  • cpe:2.3:o:philips:intellivue_mp2_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mp2:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_x2_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_x2:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mp30_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mp30:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mp50_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mp50:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mp70_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mp70:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
Show 26 more CPE
CPEVersionsRoleVendor
cpe:2.3:o:philips:intellivue_np90_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_np90:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mx700_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mx700:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mx800_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mx800:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mx400_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mx400:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mx450_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mx450:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mx500_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mx500:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mx550_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mx550:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_x3_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_x3:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:intellivue_mx100_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:intellivue_mx100:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm20_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm20:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm30_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm30:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm40_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm40:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm50_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerablePhilips
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm50:-:*:*:*:*:*:*:*not specified by NVDRunning onPhilips
  • cpe:2.3:o:philips:intellivue_np90_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_np90:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mx700_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mx700:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mx800_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mx800:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mx400_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mx400:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mx450_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mx450:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mx500_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mx500:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mx550_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mx550:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_x3_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_x3:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:intellivue_mx100_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:intellivue_mx100:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm20_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm20:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm30_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm30:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm40_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm40:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips
  • cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm50_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Philips
  • cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm50:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Philips

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2018-10601
  2. 2026-10-03 Alias added: GHSA-9f6x-r29g-cv4j

References

  • TI-2018-07486 Improper authentication in ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal… CVSS 3.1HIGH 8.3
  • TI-2018-07488 Information disclosure in ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal… CVSS 3.0MEDIUM 5.3

Sources