TI-2018-13128 Aliases CVE-2018-18564GHSA-55x2-4vmg-3vq2

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial…

Priority LOW 33Published · Updated

Key facts

Attack vector
Adjacent network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
—
Confidentiality · integrity · availability
None · High · None
Fix status
No fix listed

Attack profile from CVSS 3.0 · NVD. Fix status from CNA.

Summary

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). Improper access control allows attackers in the adjacent network to change the instrument configuration.

Why it matters now

EPSS 0.00584 48th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
——all versions—No fix listedCNA
  • — Vendor— Affected all versions Fixed in — No fix listed CNA

CPE configurations NVD · 6

CPEVersionsRoleVendor
cpe:2.3:o:roche:accu-chek_inform_ii_firmware:*:*:*:*:*:*:*:*before 03.06.00; 04.00.00 to before 04.03.00VulnerableRoche
cpe:2.3:h:roche:accu-chek_inform_ii:-:*:*:*:*:*:*:*not specified by NVDRunning onRoche
cpe:2.3:o:roche:cobas_h_232_firmware:*:*:*:*:*:*:*:*before 04.00.04VulnerableRoche
cpe:2.3:h:roche:cobas_h_232:-:*:*:*:*:*:*:*not specified by NVDRunning onRoche
cpe:2.3:o:roche:coaguchek_pro_ii_firmware:*:*:*:*:*:*:*:*before 04.03.00VulnerableRoche
cpe:2.3:h:roche:coaguchek_pro_ii:-:*:*:*:*:*:*:*not specified by NVDRunning onRoche
  • cpe:2.3:o:roche:accu-chek_inform_ii_firmware:*:*:*:*:*:*:*:* Versions before 03.06.00; 04.00.00 to before 04.03.00 Vulnerable ·Roche
  • cpe:2.3:h:roche:accu-chek_inform_ii:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Roche
  • cpe:2.3:o:roche:cobas_h_232_firmware:*:*:*:*:*:*:*:* Versions before 04.00.04 Vulnerable ·Roche
  • cpe:2.3:h:roche:cobas_h_232:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Roche
  • cpe:2.3:o:roche:coaguchek_pro_ii_firmware:*:*:*:*:*:*:*:* Versions before 04.03.00 Vulnerable ·Roche
  • cpe:2.3:h:roche:coaguchek_pro_ii:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Roche

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2018-18564
  2. 2026-10-03 Alias added: GHSA-55x2-4vmg-3vq2

References

  • TI-2018-13127 An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial… CVSS 3.0CRITICAL 9.6
  • TI-2018-13126 An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before… CVSS 3.0HIGH 8.8
  • TI-2018-13125 An issue was discovered in Roche Accu-Chek Inform II Base Unit / Base Unit Hub before… CVSS 3.0HIGH 8.0
  • TI-2018-13129 An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial… CVSS 3.0MEDIUM 6.8

Sources