An issue was discovered on Dr Trust ECG Pen 2.00.08 devices.
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Impact type
- —
- Confidentiality · integrity · availability
- High · None · None
- Fix status
- No fix listed
Attack profile from CVSS 3.1 · NVD. Fix status from CNA.
Summary
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Why it matters now
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| — | — | all versions | — | No fix listed | CNA |
- — Vendor— Affected all versions Fixed in — No fix listed CNA
Remediation
No fixed version listed by the sources yet.