Hard-coded credentials in Hamilton Medical AG, T1-Ventillator
Priority LOW 23CWE-798Published · Updated
Key facts
- Attack vector
- Physical
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Impact type
- Security bypass
- Confidentiality · integrity · availability
- High · Low · None
- Fix status
- No fix listed
Attack profile from CVSS 3.1 · NVD. Fix status from CNA. Impact type from CWE-798.
Summary
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the device’s configuration interface.
Why it matters now
EPSS 0.00282 20th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| — | Hamilton Medical AG, T1-Ventillator | up to 2.2.3 | — | No fix listed | CNA |
- Hamilton Medical AG, T1-Ventillator Vendor— Affected up to 2.2.3 Fixed in — No fix listed CNA
CPE configurations NVD · 2
| CPE | Versions | Role | Vendor |
|---|---|---|---|
| cpe:2.3:o:hamilton-medical:hamilton-t1_firmware:*:*:*:*:*:*:*:* | up to 2.2.3 | Vulnerable | Hamilton-medical |
| cpe:2.3:h:hamilton-medical:hamilton-t1:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Hamilton-medical |
- cpe:2.3:o:hamilton-medical:hamilton-t1_firmware:*:*:*:*:*:*:*:* Versions up to 2.2.3 Vulnerable ·Hamilton-medical
- cpe:2.3:h:hamilton-medical:hamilton-t1:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Hamilton-medical
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2020-27278
- 2026-10-03 Alias added: GHSA-pr6m-xr2g-5w3x
References
More Hamilton-medical Hamilton-t1 Firmware vulnerabilities
- TI-2020-16613 Information disclosure in Hamilton Medical AG, T1-Ventillator CVSS 3.1MEDIUM 4.3
- TI-2020-16606 In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation… CVSS 3.1MEDIUM 4.3
Sources
- CVE-2020-27278 · CVE Program
- CVE-2020-27278 · NVD
- FIRST EPSS
- GHSA-pr6m-xr2g-5w3x · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences