Read the docs
TI-2021-05963 Aliases CVE-2021-23195GHSA-29g2-4fxx-7xmh

Fresenius Kabi Agilia Connect Infusion System exposure of information through directory listing

Priority LOW 32CWE-548Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Information disclosure
Confidentiality · integrity · availability
Low · None · None
Fix status
Fixed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-200.

Summary

Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exist and directory listing is enabled, all content of the directory will be displayed, allowing an attacker to identify and access files on the server.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00845 58th percentileScored 2026-11-07
Fix Fixed Vigilant Software Suite (Mastermed Dashboard) (Fresenius Kabi): 2.0.1.3

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
Yes
Technical impact
Partial

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Fresenius KabiVigilant Software Suite (Mastermed Dashboard)before 2.0.1.32.0.1.3Patch availableCNA
  • Vigilant Software Suite (Mastermed Dashboard) VendorFresenius Kabi Affected before 2.0.1.3 Fixed in 2.0.1.3 Patch available CNA

CPE configurations NVD · 10

CPEVersionsRoleVendor
cpe:2.3:o:fresenius-kabi:agilia_connect_firmware:*:*:*:*:*:*:*:*up to d25VulnerableFresenius Kabi
cpe:2.3:h:fresenius-kabi:agilia_connect:-:*:*:*:*:*:*:*not specified by NVDRunning onFresenius Kabi
cpe:2.3:a:fresenius-kabi:agilia_partner_maintenance_software:*:*:*:*:*:*:*:*up to 3.3.0VulnerableFresenius Kabi
cpe:2.3:a:fresenius-kabi:vigilant_centerium:1.0:*:*:*:*:*:*:*1.0VulnerableFresenius Kabi
cpe:2.3:a:fresenius-kabi:vigilant_insight:1.0:*:*:*:*:*:*:*1.0VulnerableFresenius Kabi
cpe:2.3:a:fresenius-kabi:vigilant_mastermed:1.0:*:*:*:*:*:*:*1.0VulnerableFresenius Kabi
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:*:*:*:*:*:*:*:*before 3.0VulnerableFresenius Kabi
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:-:*:*:*:*:*:*3.0VulnerableFresenius Kabi
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:d15:*:*:*:*:*:*3.0 d15VulnerableFresenius Kabi
cpe:2.3:h:fresenius-kabi:link\+_agilia:-:*:*:*:*:*:*:*not specified by NVDRunning onFresenius Kabi
  • cpe:2.3:o:fresenius-kabi:agilia_connect_firmware:*:*:*:*:*:*:*:* Versions up to d25 Vulnerable ·Fresenius Kabi
  • cpe:2.3:h:fresenius-kabi:agilia_connect:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:agilia_partner_maintenance_software:*:*:*:*:*:*:*:* Versions up to 3.3.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:vigilant_centerium:1.0:*:*:*:*:*:*:* Versions 1.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:vigilant_insight:1.0:*:*:*:*:*:*:* Versions 1.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:vigilant_mastermed:1.0:*:*:*:*:*:*:* Versions 1.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:*:*:*:*:*:*:*:* Versions before 3.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:-:*:*:*:*:*:* Versions 3.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:d15:*:*:*:*:*:* Versions 3.0 d15 Vulnerable ·Fresenius Kabi
  • cpe:2.3:h:fresenius-kabi:link\+_agilia:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Fresenius Kabi

Remediation

  • Fix · CNA Vigilant Software Suite (Mastermed Dashboard) (Fresenius Kabi) Fixed in 2.0.1.3

Fixed in the same release

Records fixed by the same vendor advisory or release, as the CNA lists them.

  • Release ·

    Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3

    6 records Highest priority Priority MEDIUM 1 MEDIUM, 5 LOW

    Other records, highest priority first

    • TI-2021-19327 Fresenius Kabi Agilia Connect Infusion System use of client side authenticationPriority MEDIUM 50
    • TI-2021-20073 Fresenius Kabi Agilia Connect Infusion System hard coded credentialsPriority LOW 39
    • TI-2021-13191 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithmPriority LOW 33

Timeline

  1. 2026-10-02 Threatist record created from CVE-2021-23195
  2. 2026-10-03 Alias added: GHSA-29g2-4fxx-7xmh

References

  • TI-2021-19327 Fresenius Kabi Agilia Connect Infusion System use of client side authentication CVSS 3.1CRITICAL 9.8
  • TI-2021-05987 Fresenius Kabi Agilia Connect Infusion System CVSS 3.1CRITICAL 9.8
  • TI-2021-05964 Fresenius Kabi Agilia Connect Infusion System insufficiently protected credentials CVSS 3.1CRITICAL 9.8
  • TI-2021-11677 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm CVSS 3.1CRITICAL 9.1
  • TI-2021-05988 Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption CVSS 3.1HIGH 7.5
  • TI-2021-20073 Fresenius Kabi Agilia Connect Infusion System hard coded credentials CVSS 3.1HIGH 8.8
See all 8 →

Sources