Read the docs
TI-2021-05972 Aliases CVE-2021-23207GHSA-4r23-36c8-4wg2

Fresenius Kabi Agilia Connect Infusion System plaintext storage of a password

Priority LOW 28CWE-256Published · Updated

Key facts

Attack vector
Local
Privileges required
Low
User interaction
None
Attack complexity
Low
Impact type
Information disclosure
Confidentiality · integrity · availability
High · None · None
Fix status
Fixed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-256.

Summary

An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00218 12th percentileScored 2026-11-07
Fix Fixed Vigilant Software Suite (Mastermed Dashboard) (Fresenius Kabi): 2.0.1.3Agilia Partner (Fresenius Kabi): 3.0

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
No
Technical impact
Partial

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Fresenius KabiVigilant Software Suite (Mastermed Dashboard)before 2.0.1.32.0.1.3Patch availableCNA
Fresenius KabiAgilia Partnerbefore 3.03.0Patch availableCNA
  • Vigilant Software Suite (Mastermed Dashboard) VendorFresenius Kabi Affected before 2.0.1.3 Fixed in 2.0.1.3 Patch available CNA
  • Agilia Partner VendorFresenius Kabi Affected before 3.0 Fixed in 3.0 Patch available CNA

CPE configurations NVD · 10

CPEVersionsRoleVendor
cpe:2.3:o:fresenius-kabi:agilia_connect:*:*:*:*:*:*:*:*up to d25VulnerableFresenius Kabi
cpe:2.3:h:fresenius-kabi:agilia_connect:-:*:*:*:*:*:*:*not specified by NVDRunning onFresenius Kabi
cpe:2.3:a:fresenius-kabi:agilia_partner_maintenance_software:*:*:*:*:*:*:*:*up to 3.3.0VulnerableFresenius Kabi
cpe:2.3:a:fresenius-kabi:vigilant_centerium:1.0:*:*:*:*:*:*:*1.0VulnerableFresenius Kabi
cpe:2.3:a:fresenius-kabi:vigilant_insight:1.0:*:*:*:*:*:*:*1.0VulnerableFresenius Kabi
cpe:2.3:a:fresenius-kabi:vigilant_mastermed:1.0:*:*:*:*:*:*:*1.0VulnerableFresenius Kabi
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:*:*:*:*:*:*:*:*before 3.0VulnerableFresenius Kabi
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:-:*:*:*:*:*:*3.0VulnerableFresenius Kabi
cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:d15:*:*:*:*:*:*3.0 d15VulnerableFresenius Kabi
cpe:2.3:h:fresenius-kabi:link\+_agilia:-:*:*:*:*:*:*:*not specified by NVDRunning onFresenius Kabi
  • cpe:2.3:o:fresenius-kabi:agilia_connect:*:*:*:*:*:*:*:* Versions up to d25 Vulnerable ·Fresenius Kabi
  • cpe:2.3:h:fresenius-kabi:agilia_connect:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:agilia_partner_maintenance_software:*:*:*:*:*:*:*:* Versions up to 3.3.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:vigilant_centerium:1.0:*:*:*:*:*:*:* Versions 1.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:vigilant_insight:1.0:*:*:*:*:*:*:* Versions 1.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:a:fresenius-kabi:vigilant_mastermed:1.0:*:*:*:*:*:*:* Versions 1.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:*:*:*:*:*:*:*:* Versions before 3.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:-:*:*:*:*:*:* Versions 3.0 Vulnerable ·Fresenius Kabi
  • cpe:2.3:o:fresenius-kabi:link\+_agilia_firmware:3.0:d15:*:*:*:*:*:* Versions 3.0 d15 Vulnerable ·Fresenius Kabi
  • cpe:2.3:h:fresenius-kabi:link\+_agilia:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Fresenius Kabi

Remediation

  • Fix · CNA Vigilant Software Suite (Mastermed Dashboard) (Fresenius Kabi) Fixed in 2.0.1.3
  • Fix · CNA Agilia Partner (Fresenius Kabi) Fixed in 3.0

Fixed in the same release

Records fixed by the same vendor advisory or release, as the CNA lists them.

  • Release ·

    Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3

    6 records Highest priority Priority MEDIUM 1 MEDIUM, 5 LOW

    Other records, highest priority first

    • TI-2021-19327 Fresenius Kabi Agilia Connect Infusion System use of client side authenticationPriority MEDIUM 50
    • TI-2021-20073 Fresenius Kabi Agilia Connect Infusion System hard coded credentialsPriority LOW 39
    • TI-2021-13191 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithmPriority LOW 33

Timeline

  1. 2026-10-02 Threatist record created from CVE-2021-23207
  2. 2026-10-03 Alias added: GHSA-4r23-36c8-4wg2

References

  • TI-2021-18458 Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm CVSS 3.1HIGH 7.5

Sources