Read the docs
TI-2021-16000 Aliases CVE-2021-38396GHSA-phv8-rwj8-6jj2

Missing Support Integrity Check for Boston Scientific Zoom Latitude

Priority LOW 29CWE-353Published · Updated

Key facts

Attack vector
Physical
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · NVD. Fix status from CNA. Impact type from CWE-345.

Summary

The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.

Why it matters now

EPSS 0.00173 6th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
Boston ScientificZOOM LATITUDEModel 3120—No fix listedCNA
  • ZOOM LATITUDE VendorBoston Scientific Affected Model 3120 Fixed in — No fix listed CNA

CPE configurations NVD · 2

CPEVersionsRoleVendor
cpe:2.3:o:bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120_firmware:*:*:*:*:*:*:*:*all versionsVulnerableBoston Scientific
cpe:2.3:h:bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120:-:*:*:*:*:*:*:*not specified by NVDRunning onBoston Scientific
  • cpe:2.3:o:bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Boston Scientific
  • cpe:2.3:h:bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Boston Scientific

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2021-38396
  2. 2026-10-03 Alias added: GHSA-phv8-rwj8-6jj2

References

  • TI-2021-15996 Improper Access Control for Boston Scientific Zoom Latitude CVSS 3.1HIGH 7.6
  • TI-2021-16002 Reliance on Component that is not Updateable for Boston Scientific Zoom Latitude CVSS 3.1MEDIUM 6.8
  • TI-2021-16004 Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude CVSS 3.1MEDIUM 6.9
  • TI-2021-15998 Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging… CVSS 3.1MEDIUM 6.4