Read the docs
TI-2023-00840 Aliases CVE-2023-0888GHSA-9w87-qf4h-4fc6

Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi

Priority LOW 33CWE-95Published · Updated

Key facts

Attack vector
Network
Privileges required
High
User interaction
None
Attack complexity
Low
Impact type
Code execution
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · NVD. Fix status from CNA. Impact type from CWE-95.

Summary

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00604 49th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
No
Technical impact
Partial

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
B. Braun Melsungen AGSpace Battery Pack SP with Wi-Fiup to 053L000092; up to 054U000092—No fix listedCNA
  • Space Battery Pack SP with Wi-Fi VendorB. Braun Melsungen AG Affected up to 053L000092; up to 054U000092 Fixed in — No fix listed CNA

CPE configurations NVD · 3

CPEVersionsRoleVendor
cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:global:*:*:*up to 053l000092VulnerableB. Braun Melsungen AG
cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:us:*:*:*up to 054u000092VulnerableB. Braun Melsungen AG
cpe:2.3:h:bbraun:battery-pack_sp_with_wifi:-:*:*:*:*:*:*:*not specified by NVDRunning onB. Braun Melsungen AG
  • cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:global:*:*:* Versions up to 053l000092 Vulnerable ·B. Braun Melsungen AG
  • cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:us:*:*:* Versions up to 054u000092 Vulnerable ·B. Braun Melsungen AG
  • cpe:2.3:h:bbraun:battery-pack_sp_with_wifi:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·B. Braun Melsungen AG

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2023-0888
  2. 2026-10-03 Alias added: GHSA-9w87-qf4h-4fc6

References