Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials
Priority LOW 38CWE-1391Published · Updated
Key facts
- Attack vector
- Local
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Attack requirements
- None
- Impact type
- —
- Confidentiality · integrity · availability
- High · High · None
- Fix status
- No fix listed
Attack profile from CVSS 4.0 · CNA. Fix status from CNA.
Summary
A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
Why it matters now
Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00127 2nd percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Total
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2025-2229
- 2026-10-03 Alias added: GHSA-x7gm-m6fc-fx8q
References
More Philips vulnerabilities
- TI-2017-00134 The SMBv1 server in Microsoft Windows Vista SP2; Windows Server 2008 SP2 and R2 SP1… KEVCVSS 3.1HIGH 8.8
- TI-2017-00184 Microsoft Office 2007 SP3, Microsoft Office 2010 SP2, Microsoft Office 2013 SP1… KEVCVSS 3.1HIGH 7.8
- TI-2018-06405 Philips EncoreAnywhere Exposure of Sensitive Information to an Unauthorized Actor CVSS 3.1HIGH 7.5
- TI-2018-03984 Improper input validation in Philips IntelliSpace Portal CVSS 3.0CRITICAL 9.8
- TI-2018-06392 Improper input validation in Philips e-Alert Unit CVSS 3.0CRITICAL 9.8
- TI-2018-03982 Access control flaw in Philips IntelliSpace Portal CVSS 3.0CRITICAL 9.8
Sources
- CVE-2025-2229 · CVE Program
- CVE-2025-2229 · NVD
- FIRST EPSS
- GHSA-x7gm-m6fc-fx8q · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences