Eppendorf BioFlo 320 Use of hard-coded password

Priority MEDIUM 49CWE-259Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Attack requirements
None
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 4.0 · CNA. Fix status from CNA. Impact type from CWE-259.

Summary

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00498 42nd percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
Yes
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
EppendorfBioFlo 320All—No fix listedCNA
  • BioFlo 320 VendorEppendorf Affected All Fixed in — No fix listed CNA

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2026-7251
  2. 2026-10-03 Alias added: GHSA-v6xg-wr2p-xrj3
  3. 2026-10-25 EPSS moved to the below the 50th percentile

References

Sources