Mira Hormone Monitor, Mira Android App Authentication bypass by spoofing
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- Active
- Attack complexity
- Low
- Attack requirements
- None
- Impact type
- Security bypass
- Confidentiality · integrity · availability
- High · Low · None
- Fix status
- No fix listed
Attack profile from CVSS 4.0 · CNA. Fix status from CNA. Impact type from CWE-290.
Summary
The Mira Android companion app v4.5.15.4 identifies the paired Mira hormone analyzer by performing a substring match against the BLE advertisement name only, with no cryptographic peripheral authentication, MAC allowlist, or bonded-identity check. An attacker could capture live session token information and inject forged hormone measurements into the victim's cloud record and clinical trend view.
Why it matters now
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Partial
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Quanovate Tech Inc. (operating as Mira / Mira Care) | Mira Firmware | 1.7.1.47 | — | No fix listed | CNA |
| Quanovate Tech Inc. (operating as Mira / Mira Care) | Mira Android App | 4.5.15.4 | — | No fix listed | CNA |
- Mira Firmware VendorQuanovate Tech Inc. (operating as Mira / Mira Care) Affected 1.7.1.47 Fixed in — No fix listed CNA
- Mira Android App VendorQuanovate Tech Inc. (operating as Mira / Mira Care) Affected 4.5.15.4 Fixed in — No fix listed CNA
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2026-67558
- 2026-10-03 Alias added: GHSA-cxxh-6vq9-57gc
References
More Quanovate Tech Inc. (operating as Mira / Mira Care) vulnerabilities
- TI-2026-47851 Mira Hormone Monitor, Mira Android App Weak Authentication CVSS 4.0CRITICAL 9.3
- TI-2026-47743 Mira Hormone Monitor, Mira Android App Use of Hard-coded Credentials CVSS 4.0CRITICAL 9.3
- TI-2026-47465 Mira Hormone Monitor, Mira Android App Missing authentication for critical function CVSS 4.0HIGH 8.7
- TI-2026-47127 Mira Hormone Monitor, Mira Android App Improper restriction of excessive authentication attempts CVSS 4.0MEDIUM 6.9
- TI-2026-47458 Mira Hormone Monitor, Mira Android App Use of GET request method with sensitive query strings CVSS 4.0MEDIUM 6.9
- TI-2026-47059 Mira Hormone Monitor, Mira Android App Missing authentication for critical function CVSS 4.0HIGH 7.1
Sources
- CVE-2026-67558 · CVE Program
- CVE-2026-67558 · NVD
- FIRST EPSS
- GHSA-cxxh-6vq9-57gc · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences