Priority MEDIUM 52
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
CVSS 2.0HIGH 10.0· Network · No auth · Interaction — · EPSS 0.06622 · Published 2015-07-06 · No fix listed CWE-119
Priority MEDIUM 51
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified com…
CVSS 2.0HIGH 10.0· Network · No auth · Interaction — · EPSS 0.05064 · Published 2015-04-29 · No fix listed CWE-264
Priority MEDIUM 50
Hospira MedNet software version 5.8 and prior uses vulnerable versions of the JBoss Enterprise Application Platform software that may allow unauthenticated users to execute arbitrary code on the target system.
CVSS 3.0CRITICAL 9.8· Network · No auth · No interaction · EPSS 0.05009 · Published 2019-03-26 · Fixed in 6.1 CWE-94
Priority MEDIUM 46
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by…
CVSS 3.0CRITICAL 9.8· Network · No auth · No interaction · EPSS 0.02036 · Published 2019-03-25 · No fix listed CWE-285
Priority MEDIUM 46
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior.
CVSS 3.0CRITICAL 9.8· Network · No auth · No interaction · EPSS 0.01972 · Published 2019-03-25 · No fix listed CWE-259
Priority MEDIUM 44
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unau…
CVSS 3.0CRITICAL 9.8· Network · No auth · No interaction · EPSS 0.00954 · Published 2019-03-25 · No fix listed CWE-345
Priority MEDIUM 43
Hospira Symbiq Infusion System 3.13 and earlier allows remote authenticated users to trigger "unanticipated operations" by leveraging "elevated privileges" for an unspecified call to an incorrectly exposed function.
CVSS 3.0HIGH 8.8· Network · Low privileges · No interaction · EPSS 0.02882 · Published 2019-03-23 · No fix listed CWE-264
Priority MEDIUM 43
Hospira MedNet before 6.1 uses a hardcoded cleartext password to control SQL database authorization, which allows remote authenticated users to bypass intended access restrictions by leveraging knowledge of this password…
CVSS 2.0CRITICAL 9.0· Network · Low privileges · Interaction — · EPSS 0.02327 · Published 2015-04-03 · Fixed in 6.1 CWE-259
Priority MEDIUM 42
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify…
CVSS 2.0HIGH 9.3· Network · No auth · Interaction — · EPSS 0.01218 · Published 2015-07-06 · Fixed in 7.0 CWE-345
Priority LOW 39
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.
CVSS 2.0HIGH 7.8· Network · No auth · Interaction — · EPSS 0.02562 · Published 2015-07-06 · No fix listed CWE-19
Priority LOW 35
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior.
CVSS 3.0HIGH 7.5· Network · No auth · No interaction · EPSS 0.01008 · Published 2019-03-25 · No fix listed CWE-312
Priority LOW 34
Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40 allows remote attackers to cause a denia…
CVSS 3.0HIGH 7.3· Network · No auth · No interaction · EPSS 0.01112 · Published 2016-01-22 · No fix listed CWE-119
Priority LOW 34
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System.
CVSS 3.0HIGH 7.5· Network · No auth · No interaction · EPSS 0.00771 · Published 2019-03-25 · No fix listed CWE-312
Priority LOW 34
Hospira MedNet before 6.1 uses hardcoded cryptographic keys for protection of data transmission from infusion pumps, which allows remote attackers to obtain sensitive information by sniffing the network.
CVSS 2.0MEDIUM 6.8· Network · Low privileges · Interaction — · EPSS 0.01744 · Published 2015-04-03 · Fixed in 6.1 CWE-321
Priority LOW 30
The installation component in Hospira MedNet before 6.1 places cleartext credentials in configuration files, which allows local users to obtain sensitive information by reading a file.
CVSS 2.0MEDIUM 6.8· Local · Low privileges · Interaction — · EPSS 0.00329 · Published 2015-04-03 · Fixed in 6.1 CWE-260
Priority LOW 27
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.
CVSS 2.0MEDIUM 5.0· Network · No auth · Interaction — · EPSS 0.02185 · Published 2015-07-06 · No fix listed CWE-200
Priority LOW 22
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.
CVSS 2.0MEDIUM 4.6· Local · No auth · Interaction — · EPSS 0.00544 · Published 2015-07-06 · No fix listed CWE-255