Read the docs
TI-2015-03342 Aliases CVE-2015-3956GHSA-2vhp-v3rh-j2qr

Insufficient verification of data authenticity in Hospira Plum A+ Infusion System

Priority MEDIUM 44CWE-345Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.0 · NVD. Fix status from CNA. Impact type from CWE-345.

Summary

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Why it matters now

EPSS 0.00954 61st percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
HospiraPlum A+ Infusion Systemup to 13.4—No fix listedCNA
HospiraPlum A+3 Infusion Systemup to 13.6—No fix listedCNA
HospiraSymbiq Infusion Systemup to 3.13—No fix listedCNA
  • Plum A+ Infusion System VendorHospira Affected up to 13.4 Fixed in — No fix listed CNA
  • Plum A+3 Infusion System VendorHospira Affected up to 13.6 Fixed in — No fix listed CNA
  • Symbiq Infusion System VendorHospira Affected up to 3.13 Fixed in — No fix listed CNA

CPE configurations NVD · 6

CPEVersionsRoleVendor
cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:*up to 13.4VulnerablePifzer
cpe:2.3:h:pifzer:plum_a\+_infusion_system:-:*:*:*:*:*:*:*not specified by NVDRunning onPifzer
cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:*up to 13.6VulnerablePifzer
cpe:2.3:h:pifzer:plum_a\+3_infusion_system:-:*:*:*:*:*:*:*not specified by NVDRunning onPifzer
cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*up to 3.13VulnerablePifzer
cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*not specified by NVDRunning onPifzer
  • cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:* Versions up to 13.4 Vulnerable ·Pifzer
  • cpe:2.3:h:pifzer:plum_a\+_infusion_system:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Pifzer
  • cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:* Versions up to 13.6 Vulnerable ·Pifzer
  • cpe:2.3:h:pifzer:plum_a\+3_infusion_system:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Pifzer
  • cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:* Versions up to 3.13 Vulnerable ·Pifzer
  • cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Pifzer

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2015-3956
  2. 2026-10-03 Alias added: GHSA-2vhp-v3rh-j2qr

References

  • TI-2015-03341 Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and… CVSS 2.0HIGH 10.0
  • TI-2015-03050 The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not… CVSS 2.0HIGH 10.0
  • TI-2014-04403 Hospira MedNet Code Injection CVSS 3.0CRITICAL 9.8
  • TI-2015-03340 Improper authorization in Hospira Plum A+ Infusion System CVSS 3.0CRITICAL 9.8
  • TI-2015-03339 Hard-coded credentials in Hospira Plum A+ Infusion System CVSS 3.0CRITICAL 9.8
  • TI-2015-03351 Access control flaw in Hospira Symbiq Infusion System CVSS 3.0HIGH 8.8
See all 16 →