Improper handling of exceptional conditions in Philips IntelliVue MX40 Patient Worn…
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Impact type
- —
- Confidentiality · integrity · availability
- None · None · High
- Fix status
- No fix listed
Attack profile from CVSS 3.0 · NVD. Fix status from CNA.
Summary
Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Why it matters now
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Philips | IntelliVue MX40 Patient Worn Monitor | IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18 | — | No fix listed | CNA |
- IntelliVue MX40 Patient Worn Monitor VendorPhilips Affected IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18 Fixed in — No fix listed CNA
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2017-9658
- 2026-10-03 Alias added: GHSA-ccgj-ph26-hf87
References
More Philips Intellivue Mx40 Firmware vulnerabilities
- TI-2017-07054 Improper handling of exceptional conditions in Philips IntelliVue MX40 Patient Worn… CVSS 3.0MEDIUM 6.5
Sources
- CVE-2017-9658 · CVE Program
- CVE-2017-9658 · NVD
- FIRST EPSS
- GHSA-ccgj-ph26-hf87 · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences