Blipcare Wifi blood pressure monitor BP700 10.1 devices allow memory corruption that…
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Impact type
- Denial of service
- Confidentiality · integrity · availability
- None · None · High
- Fix status
- No fix listed
Attack profile from CVSS 3.0 · NVD. Fix status from CNA. Impact type from CWE-399 and the CVSS impact.
Summary
Blipcare Wifi blood pressure monitor BP700 10.1 devices allow memory corruption that results in Denial of Service. When connected to the "Blip" open wireless connection provided by the device, if a large string is sent as a part of the HTTP request in any part of the HTTP headers, the device could become completely unresponsive. Presumably this happens as the memory footprint provided to this device is very small. According to the specs from Rezolt, the Wi-Fi module only has 256k of memory. As a result, an incorrect string copy operation using either memcpy, strcpy, or any of their other variants could result in filling up the memory space allocated to the function executing and this would result in memory corruption. To test the theory, one can modify the demo application provided by the Cypress WICED SDK and introduce an incorrect "memcpy" operation and use the compiled application on the evaluation board provided by Cypress semiconductors with exactly the same Wi-Fi SOC. The results were identical where the device would completely stop responding to any of the ping or web requests.
Why it matters now
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| — | — | all versions | — | No fix listed | CNA |
- — Vendor— Affected all versions Fixed in — No fix listed CNA
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2017-11580
- 2026-10-03 Alias added: GHSA-857g-5g59-pwmx
References
More Blipcare Wi-fi Blood Pressure Monitor Firmware vulnerabilities
- TI-2017-08573 In the most recent firmware for Blipcare, the device provides an open Wireless network… CVSS 3.0HIGH 7.1
- TI-2017-08572 It was discovered as a part of the research on IoT devices in the most recent firmware… CVSS 3.0MEDIUM 5.9
Sources
- CVE-2017-11580 · CVE Program
- CVE-2017-11580 · NVD
- FIRST EPSS
- GHSA-857g-5g59-pwmx · GitHub Advisory Database · CC BY 4.0