Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
Key facts
- Attack vector
- Physical
- Privileges required
- Low
- User interaction
- None
- Attack complexity
- High
- Impact type
- —
- Confidentiality · integrity · availability
- High · Low · Low
- Fix status
- No fix listed
Attack profile from CVSS 3.1 · CNA. Fix status from CNA.
Summary
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Why it matters now
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Medtronic | 24950 MyCareLink Monitor | All versions | — | No fix listed | CNA |
| Medtronic | 24952 MyCareLink Monitor | All versions | — | No fix listed | CNA |
- 24950 MyCareLink Monitor VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- 24952 MyCareLink Monitor VendorMedtronic Affected All versions Fixed in — No fix listed CNA
CPE configurations NVD · 4
| CPE | Versions | Role | Vendor |
|---|---|---|---|
| cpe:2.3:o:medtronic:24950_mycarelink_monitor_firmware:-:*:*:*:*:*:*:* | not specified by NVD | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:24950_mycarelink_monitor:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:24952_mycarelink_monitor_firmware:-:*:*:*:*:*:*:* | not specified by NVD | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:24952_mycarelink_monitor:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
- cpe:2.3:o:medtronic:24950_mycarelink_monitor_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:24950_mycarelink_monitor:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:24952_mycarelink_monitor_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:24952_mycarelink_monitor:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2018-8868
- 2026-10-03 Alias added: GHSA-c9h8-j583-w6wh
References
More Medtronic 24950 Mycarelink Monitor Firmware vulnerabilities
- TI-2018-06412 Medtronic MyCareLink Patient Monitor Use of Hard-coded Password CVSS 3.1MEDIUM 6.4
Sources
- CVE-2018-8868 · CVE Program
- CVE-2018-8868 · NVD
- FIRST EPSS
- GHSA-c9h8-j583-w6wh · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences