Read the docs
TI-2018-07485 Aliases CVE-2018-10596GHSA-3vcw-xhqc-97mh

Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints

Priority LOW 34CWE-923Published · Updated

Key facts

Attack vector
Network
Privileges required
Low
User interaction
Required
Attack complexity
High
Impact type
Information disclosure
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-200.

Summary

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

Why it matters now

EPSS 0.0132 71st percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
Medtronic2090 CareLink ProgrammerAll versions—No fix listedCNA
Medtronic29901 Encore ProgrammerAll versions—No fix listedCNA
  • 2090 CareLink Programmer VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • 29901 Encore Programmer VendorMedtronic Affected All versions Fixed in — No fix listed CNA

CPE configurations NVD · 2

CPEVersionsRoleVendor
cpe:2.3:o:medtronic:2090_carelink_programmer_firmware:*:*:*:*:*:*:*:*all versionsVulnerableMedtronic
cpe:2.3:h:medtronic:2090_carelink_programmer:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronic
  • cpe:2.3:o:medtronic:2090_carelink_programmer_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
  • cpe:2.3:h:medtronic:2090_carelink_programmer:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2018-10596
  2. 2026-10-03 Alias added: GHSA-3vcw-xhqc-97mh

References

  • TI-2018-03961 Medtronic 2090 Carelink Programmer Relative Path Traversal CVSS 3.1MEDIUM 4.8
  • TI-2018-03959 Medtronic 2090 Carelink Programmer Storing Passwords in a Recoverable Format CVSS 3.1MEDIUM 4.9

Sources