Read the docs
TI-2018-07518 Aliases CVE-2018-10631GHSA-4g99-xhm8-2rwc

Medtronic N'Vision Clinician Programmer Protection Mechanism Failure

Priority LOW 30CWE-693Published · Updated

Key facts

Attack vector
Physical
Privileges required
None
User interaction
Required
Attack complexity
High
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-693.

Summary

The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00411 35th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
No
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Medtronic8840 N’Vision Clinician ProgrammerAll versions—No fix listedCNA
Medtronic8870 N’Vision removable Application CardAll versions—No fix listedCNA
  • 8840 N’Vision Clinician Programmer VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • 8870 N’Vision removable Application Card VendorMedtronic Affected All versions Fixed in — No fix listed CNA

CPE configurations NVD · 4

CPEVersionsRoleVendor
cpe:2.3:o:medtronic:n\'vision_8840_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronic
cpe:2.3:h:medtronic:n\'vision_8840:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronic
cpe:2.3:o:medtronic:n\'vision_8870_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronic
cpe:2.3:h:medtronic:n\'vision_8870:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronic
  • cpe:2.3:o:medtronic:n\'vision_8840_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronic
  • cpe:2.3:h:medtronic:n\'vision_8840:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
  • cpe:2.3:o:medtronic:n\'vision_8870_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronic
  • cpe:2.3:h:medtronic:n\'vision_8870:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2018-10631
  2. 2026-10-03 Alias added: GHSA-4g99-xhm8-2rwc

References

  • TI-2018-06391 Medtronic N'Vision Clinician Programmer Missing Encryption of Sensitive Data CVSS 3.1MEDIUM 4.6

Sources