Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Priority LOW 25CWE-294Published · Updated

Key facts

Attack vector
Adjacent network
Privileges required
None
User interaction
None
Attack complexity
High
Impact type
Security bypass
Confidentiality · integrity · availability
None · High · None
Fix status
No fix listed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-294.

Summary

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Why it matters now

EPSS 0.00712 53rd percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
MedtronicMMT- 508 - MiniMed pumpAll versions—No fix listedCNA
MedtronicMMT – 511 pump ParadigmAll versions—No fix listedCNA
MedtronicMMT – 512 / MMT – 712 Paradigm x12All versions—No fix listedCNA
MedtronicMMT – 515 / MMT – 715 Paradigm x15All versions—No fix listedCNA
MedtronicMMT – 522 / MMT – 722 Paradigm REAL-TIMEAll versions—No fix listedCNA
MedtronicMMT – 522(K) / MMT – 722(K) Paradigm REAL-TIMEAll versions—No fix listedCNA
MedtronicMMT – 523 / MMT – 723 Paradigm RevelAll versions—No fix listedCNA
MedtronicMMT – 523(K) / MMT – 723(K) ParadigmAll versions—No fix listedCNA
MedtronicMMT – 554 / MMT – 754 MiniMed VeoAll versions—No fix listedCNA
MedtronicMMT – 551 / MMT – 751 MiniMed 530GAll versions—No fix listedCNA
  • MMT- 508 - MiniMed pump VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 511 pump Paradigm VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 512 / MMT – 712 Paradigm x12 VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 515 / MMT – 715 Paradigm x15 VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 522 / MMT – 722 Paradigm REAL-TIME VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 523 / MMT – 723 Paradigm Revel VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 523(K) / MMT – 723(K) Paradigm VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 554 / MMT – 754 MiniMed Veo VendorMedtronic Affected All versions Fixed in — No fix listed CNA
  • MMT – 551 / MMT – 751 MiniMed 530G VendorMedtronic Affected All versions Fixed in — No fix listed CNA

CPE configurations NVD · 18

CPEVersionsRoleVendor
cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
Show 8 more CPE
CPEVersionsRoleVendor
cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:551_minimed_530g:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableMedtronicdiabetes
cpe:2.3:h:medtronicdiabetes:751_minimed_530g:-:*:*:*:*:*:*:*not specified by NVDRunning onMedtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:551_minimed_530g:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes
  • cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronicdiabetes
  • cpe:2.3:h:medtronicdiabetes:751_minimed_530g:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronicdiabetes

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2018-14781
  2. 2026-10-03 Alias added: GHSA-cv4q-4xgp-cr7m

References

  • TI-2023-14928 Medtronic Paceart MSMQ Deserialization of Untrusted Data CVSS 3.1CRITICAL 9.8
  • TI-2020-15429 Medtronic MyCareLink Smart Heap-based Buffer Overflow CVSS 3.1CRITICAL 9.8
  • TI-2020-16578 Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition CVSS 3.1HIGH 8.8
  • TI-2025-11897 Missing brute-force protection in Medtronic CareLink Network CVSS 3.1CRITICAL 9.8
  • TI-2019-04592 Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control CVSS 3.1CRITICAL 9.3
  • TI-2020-15425 Medtronic MyCareLink Smart Improper Authentication CVSS 3.1HIGH 8.8
See all 34 →

Sources