Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- High
- Impact type
- Security bypass
- Confidentiality · integrity · availability
- Low · High · High
- Fix status
- No fix listed
Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-284.
Summary
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Why it matters now
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Medtronic | MiniMed 508 pump | All versions | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 511 pump | All versions | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 512/712 pumps | All versions | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 712E pump | All versions | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 515/715 pumps | All versions | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 522/722 pumps | All versions | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 522K/722K pumps | All versions | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 523/723 pumps | up to Software Versions 2.4A | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm 523K/723K pumps | up to Software Versions 2.4A | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm Veo 554/754 pumps | up to Software Versions 2.6A | — | No fix listed | CNA |
| Medtronic | MiniMed Paradigm Veo 554CM/754CM pumps | up to Software Versions 2.7A | — | No fix listed | CNA |
- MiniMed 508 pump VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- MiniMed Paradigm 511 pump VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- MiniMed Paradigm 512/712 pumps VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- MiniMed Paradigm 712E pump VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- MiniMed Paradigm 515/715 pumps VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- MiniMed Paradigm 522/722 pumps VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- MiniMed Paradigm 522K/722K pumps VendorMedtronic Affected All versions Fixed in — No fix listed CNA
- MiniMed Paradigm 523/723 pumps VendorMedtronic Affected up to Software Versions 2.4A Fixed in — No fix listed CNA
- MiniMed Paradigm 523K/723K pumps VendorMedtronic Affected up to Software Versions 2.4A Fixed in — No fix listed CNA
- MiniMed Paradigm Veo 554/754 pumps VendorMedtronic Affected up to Software Versions 2.6A Fixed in — No fix listed CNA
- MiniMed Paradigm Veo 554CM/754CM pumps VendorMedtronic Affected up to Software Versions 2.7A Fixed in — No fix listed CNA
CPE configurations NVD · 38
| CPE | Versions | Role | Vendor |
|---|---|---|---|
| cpe:2.3:o:medtronic:minimed_508_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_508:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_511_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_511:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_512_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_512:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_712_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_712:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_712e_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_712e:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
- cpe:2.3:o:medtronic:minimed_508_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_508:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_511_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_511:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_512_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_512:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_712_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_712:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_712e_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_712e:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
Show 28 more CPE
| CPE | Versions | Role | Vendor |
|---|---|---|---|
| cpe:2.3:o:medtronic:minimed_paradigm_515_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_515:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_715_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_715:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_522_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_522:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_722_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_722:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_522k_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_522k:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_722k_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_722k:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_523_firmware:*:*:*:*:*:*:*:* | up to 2.4a | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_523:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_723_firmware:*:*:*:*:*:*:*:* | up to 2.4a | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_723:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_523k_firmware:*:*:*:*:*:*:*:* | up to 2.4a | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_523k:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_723k_firmware:*:*:*:*:*:*:*:* | up to 2.4a | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_723k:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_veo_554_firmware:*:*:*:*:*:*:*:* | up to 2.6a | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_veo_554:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_veo_754_firmware:*:*:*:*:*:*:*:* | up to 2.6a | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_veo_754:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_veo_554cm_firmware:*:*:*:*:*:*:*:* | up to 2.7a | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_veo_554cm:*:*:*:*:*:*:*:* | up to 2.7a | Running on | Medtronic |
| cpe:2.3:o:medtronic:minimed_paradigm_veo_754cm_firmware:-:*:*:*:*:*:*:* | not specified by NVD | Vulnerable | Medtronic |
| cpe:2.3:h:medtronic:minimed_paradigm_veo_754cm:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Medtronic |
- cpe:2.3:o:medtronic:minimed_paradigm_515_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_515:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_715_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_715:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_522_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_522:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_722_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_722:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_522k_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_522k:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_722k_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_722k:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_523_firmware:*:*:*:*:*:*:*:* Versions up to 2.4a Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_523:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_723_firmware:*:*:*:*:*:*:*:* Versions up to 2.4a Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_723:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_523k_firmware:*:*:*:*:*:*:*:* Versions up to 2.4a Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_523k:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_723k_firmware:*:*:*:*:*:*:*:* Versions up to 2.4a Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_723k:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_veo_554_firmware:*:*:*:*:*:*:*:* Versions up to 2.6a Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_veo_554:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_veo_754_firmware:*:*:*:*:*:*:*:* Versions up to 2.6a Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_veo_754:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_veo_554cm_firmware:*:*:*:*:*:*:*:* Versions up to 2.7a Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_veo_554cm:*:*:*:*:*:*:*:* Versions up to 2.7a Running on ·Medtronic
- cpe:2.3:o:medtronic:minimed_paradigm_veo_754cm_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Medtronic
- cpe:2.3:h:medtronic:minimed_paradigm_veo_754cm:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Medtronic
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2019-10964
- 2026-10-03 Alias added: GHSA-54f2-q9g6-3cvc
References
Sources
- CVE-2019-10964 · CVE Program
- CVE-2019-10964 · NVD
- FIRST EPSS
- GHSA-54f2-q9g6-3cvc · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences