Read the docs
TI-2019-13269 Aliases CVE-2019-18252GHSA-2f8f-gw52-m98c

Improper authentication in BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM

Priority LOW 21CWE-287Published · Updated

Key facts

Attack vector
Adjacent network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
Low · None · None
Fix status
No fix listed

Attack profile from CVSS 3.1 · NVD. Fix status from CNA. Impact type from CWE-287.

Summary

BIOTRONIK CardioMessenger II, The affected products allow credential reuse for multiple authentication purposes. An attacker with adjacent access to the CardioMessenger can disclose its credentials used for connecting to the BIOTRONIK Remote Communication infrastructure.

Why it matters now

EPSS 0.00484 41st percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
—BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSMCardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20—No fix listedCNA
  • BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM Vendor— Affected CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20 Fixed in — No fix listed CNA

CPE configurations NVD · 4

CPEVersionsRoleVendor
cpe:2.3:o:biotronik:cardiomessenger_ii-s_gsm_firmware:2.20:*:*:*:*:*:*:*2.20VulnerableBiotronik
cpe:2.3:h:biotronik:cardiomessenger_ii-s_gsm:-:*:*:*:*:*:*:*not specified by NVDRunning onBiotronik
cpe:2.3:o:biotronik:cardiomessenger_ii-s_t-line_firmware:2.20:*:*:*:*:*:*:*2.20VulnerableBiotronik
cpe:2.3:h:biotronik:cardiomessenger_ii-s_t-line:-:*:*:*:*:*:*:*not specified by NVDRunning onBiotronik
  • cpe:2.3:o:biotronik:cardiomessenger_ii-s_gsm_firmware:2.20:*:*:*:*:*:*:* Versions 2.20 Vulnerable ·Biotronik
  • cpe:2.3:h:biotronik:cardiomessenger_ii-s_gsm:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Biotronik
  • cpe:2.3:o:biotronik:cardiomessenger_ii-s_t-line_firmware:2.20:*:*:*:*:*:*:* Versions 2.20 Vulnerable ·Biotronik
  • cpe:2.3:h:biotronik:cardiomessenger_ii-s_t-line:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Biotronik

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2019-18252
  2. 2026-10-03 Alias added: GHSA-2f8f-gw52-m98c

References

  • TI-2019-13273 Insufficiently protected credentials in BIOTRONIK CardioMessenger II-S T-Line… CVSS 3.1MEDIUM 4.6
  • TI-2019-13271 Cleartext storage of sensitive information in BIOTRONIK CardioMessenger II-S T-Line… CVSS 3.1MEDIUM 4.6
  • TI-2019-13263 Improper authentication in BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM CVSS 3.1MEDIUM 4.3
  • TI-2019-13265 Cleartext transmission of sensitive information in BIOTRONIK CardioMessenger II-S T-Line… CVSS 3.1MEDIUM 4.3

Sources