Read the docs
TI-2019-13273 Aliases CVE-2019-18256GHSA-qhvv-89m3-h89m

Insufficiently protected credentials in BIOTRONIK CardioMessenger II-S T-Line…

Priority LOW 21CWE-257Published · Updated

Key facts

Attack vector
Physical
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Information disclosure
Confidentiality · integrity · availability
High · None · None
Fix status
No fix listed

Attack profile from CVSS 3.1 · NVD. Fix status from CNA. Impact type from CWE-522.

Summary

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Why it matters now

EPSS 0.00358 29th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
—BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSMCardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20—No fix listedCNA
  • BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM Vendor— Affected CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20 Fixed in — No fix listed CNA

CPE configurations NVD · 4

CPEVersionsRoleVendor
cpe:2.3:o:biotronik:cardiomessenger_ii-s_gsm_firmware:2.20:*:*:*:*:*:*:*2.20VulnerableBiotronik
cpe:2.3:h:biotronik:cardiomessenger_ii-s_gsm:-:*:*:*:*:*:*:*not specified by NVDRunning onBiotronik
cpe:2.3:o:biotronik:cardiomessenger_ii-s_t-line_firmware:2.20:*:*:*:*:*:*:*2.20VulnerableBiotronik
cpe:2.3:h:biotronik:cardiomessenger_ii-s_t-line:-:*:*:*:*:*:*:*not specified by NVDRunning onBiotronik
  • cpe:2.3:o:biotronik:cardiomessenger_ii-s_gsm_firmware:2.20:*:*:*:*:*:*:* Versions 2.20 Vulnerable ·Biotronik
  • cpe:2.3:h:biotronik:cardiomessenger_ii-s_gsm:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Biotronik
  • cpe:2.3:o:biotronik:cardiomessenger_ii-s_t-line_firmware:2.20:*:*:*:*:*:*:* Versions 2.20 Vulnerable ·Biotronik
  • cpe:2.3:h:biotronik:cardiomessenger_ii-s_t-line:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Biotronik

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2019-18256
  2. 2026-10-03 Alias added: GHSA-qhvv-89m3-h89m

References

  • TI-2019-13271 Cleartext storage of sensitive information in BIOTRONIK CardioMessenger II-S T-Line… CVSS 3.1MEDIUM 4.6
  • TI-2019-13269 Improper authentication in BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM CVSS 3.1MEDIUM 4.3
  • TI-2019-13263 Improper authentication in BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM CVSS 3.1MEDIUM 4.3
  • TI-2019-13265 Cleartext transmission of sensitive information in BIOTRONIK CardioMessenger II-S T-Line… CVSS 3.1MEDIUM 4.3

Sources