Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

Priority MEDIUM 43CWE-924Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Attack requirements
None
Impact type
—
Confidentiality · integrity · availability
Low · Low · High
Fix status
Fixed

Attack profile from CVSS 4.0 · CNA. Fix status from CNA.

Summary

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00132 2nd percentileScored 2026-11-07
Fix Fixed Infinity Acute Care System (Dräger): VG4.2Standalone Infinity M540 patient monitor (Dräger): VG4.2

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
Yes
Technical impact
Partial

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
DrägerInfinity Acute Care Systemup to VG4.1.1; up to VG4.0.3VG4.2Patch availableCNA
DrägerStandalone Infinity M540 patient monitorup to VG4.1.1VG4.2Patch availableCNA
  • Infinity Acute Care System VendorDräger Affected up to VG4.1.1; up to VG4.0.3 Fixed in VG4.2 Patch available CNA
  • Standalone Infinity M540 patient monitor VendorDräger Affected up to VG4.1.1 Fixed in VG4.2 Patch available CNA

Remediation

  • Fix · CNA Infinity Acute Care System (Dräger) Fixed in VG4.2
  • Fix · CNA Standalone Infinity M540 patient monitor (Dräger) Fixed in VG4.2

Timeline

  1. 2026-10-02 Threatist record created from CVE-2019-25719
  2. 2026-10-03 Alias added: GHSA-hxqp-hf33-px7g

References

  • TI-2021-09347 Draeger X-Dock Firmware before 03.00.13 has Hard-Coded Credentials, leading to remote… CVSS 3.1HIGH 8.8
  • TI-2024-13029 Dräger Core 1.0.5 Denial of Service via Malformed SDC Message CVSS 4.0HIGH 8.7
  • TI-2022-04579 Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering CVSS 4.0HIGH 8.8
  • TI-2021-09348 Draeger X-Dock Firmware before 03.00.13 has Active Debug Code on a debug port, leading to… CVSS 3.1HIGH 8.8
  • TI-2019-15769 Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass CVSS 4.0HIGH 8.6
  • TI-2025-27661 Draeger: ICMHelper is vulnerable to a privilege escalation due too missing authorization CVSS 3.1HIGH 7.8
See all 22 →

Sources