Read the docs
TI-2024-09704 Aliases CVE-2024-10476GHSA-2qgh-37rm-mv6x

Default credentials are used in the above listed BD Diagnostic Solutions products.

Priority LOW 36CWE-1392Published · Updated

Key facts

Attack vector
Adjacent network
Privileges required
Low
User interaction
None
Attack complexity
Low
Impact type
—
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA.

Summary

Default credentials are used in the above listed BD Diagnostic Solutions products. If exploited, threat actors may be able to access, modify or delete data, including sensitive information such as protected health information (PHI) and personally identifiable information (PII). Exploitation of this vulnerability may allow an attacker to shut down or otherwise impact the availability of the system. Note: BD Synapsys™ Informatics Solution is only in scope of this vulnerability when installed on a NUC server. BD Synapsys™ Informatics Solution installed on a customer-provided virtual machine or on the BD Kiestra™ SCU hardware is not in scope.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00236 14th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
No
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Becton, Dickinson and Company (BD)BD BACTEC™ Blood Culture Systemup to 7.20—No fix listedCNA
Becton, Dickinson and Company (BD)BD COR™ Systemup to 8.90—No fix listedCNA
Becton, Dickinson and Company (BD)BD EpiCenter™ Microbiology Data Management Systemup to 7.45—No fix listedCNA
Becton, Dickinson and Company (BD)BD MAX™ Systemup to 6.10—No fix listedCNA
Becton, Dickinson and Company (BD)BD Phoenix™ M50 Automated Microbiology Systemup to 2.70—No fix listedCNA
Becton, Dickinson and Company (BD)BD Synapsys™ Informatics Solutionup to 6.10—No fix listedCNA

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2024-10476
  2. 2026-10-03 Alias added: GHSA-2qgh-37rm-mv6x

References

  • TI-2019-08178 Unrestricted file upload in BD Alaris Gateway Workstation CVSS 3.0CRITICAL 10.0
  • TI-2018-10616 Improper authentication in Becton, Dickinson and Company (BD) Alaris GS, Alaris GH… CVSS 3.0CRITICAL 9.4
  • TI-2017-03932 Hard-coded credentials in BD Kiestra PerformA and KLA Journal Service CVSS 3.0CRITICAL 9.8
  • TI-2019-10149 Session fixation in BD CVSS 3.0HIGH 8.8
  • TI-2022-07017 BD Pyxis™ Products – Default Credentials CVSS 3.1HIGH 8.8
  • TI-2023-14398 Stored Cross-Site Scripting on User Import Functionality CVSS 3.1HIGH 8.2
See all 33 →

Sources