Privilege boundary violation in Radiometer Products
Key facts
- Attack vector
- Physical
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Impact type
- Security bypass
- Confidentiality · integrity · availability
- High · High · High
- Fix status
- No fix listed
Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-284.
Summary
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
Why it matters now
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Partial
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | All application software versions with Windows 7, Windows XP as underlying OS; Application software versions < 3.5MR11 with Windows 10 as underlying OS | — | No fix listed | CNA |
| Radiometer Medical Aps | AQT90 FLEX Analyzers | All Application software versions <= 8.13 MR2 | — | No fix listed | CNA |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX Analyzers | Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS; Application software versions < 6.20 MR2 with Windows 10 as underlying OS | — | No fix listed | CNA |
| Radiometer Medical Aps | ABL9 Analyzers | Application software versions < 1.5.0 | — | No fix listed | CNA |
- ABL90 FLEX and ABL90 FLEX PLUS Analyzers VendorRadiometer Medical Aps Affected All application software versions with Windows 7, Windows XP as underlying OS; Application software versions < 3.5MR11 with Windows 10 as underlying OS Fixed in — No fix listed CNA
- AQT90 FLEX Analyzers VendorRadiometer Medical Aps Affected All Application software versions <= 8.13 MR2 Fixed in — No fix listed CNA
- ABL800 BASIC and ABL800 FLEX Analyzers VendorRadiometer Medical Aps Affected Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS; Application software versions < 6.20 MR2 with Windows 10 as underlying OS Fixed in — No fix listed CNA
- ABL9 Analyzers VendorRadiometer Medical Aps Affected Application software versions < 1.5.0 Fixed in — No fix listed CNA
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2025-14095
- 2026-10-03 Alias added: GHSA-jvwx-cqx2-8qq4
References
More Radiometer Medical Aps vulnerabilities
- TI-2025-12881 Credential Disclosure vulnerability in Radiometer Products CVSS 3.1HIGH 8.4
- TI-2025-12882 Remote Code Execution Vulnerability in Radiometer Products CVSS 3.1HIGH 7.2
Sources
- CVE-2025-14095 · CVE Program
- CVE-2025-14095 · NVD
- FIRST EPSS
- GHSA-jvwx-cqx2-8qq4 · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences