Remote Code Execution Vulnerability in Radiometer Products
Key facts
- Attack vector
- Network
- Privileges required
- High
- User interaction
- None
- Attack complexity
- Low
- Impact type
- Security bypass
- Confidentiality · integrity · availability
- High · High · High
- Fix status
- Fixed
Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-287.
Summary
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Why it matters now
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Total
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems | Application software versions >= 3.5MR11 with Windows 10 operating system | Patch available | CNA |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX Analyzers | Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems | Application software versions >= 6.20MR2 with Windows 10 operating system | Patch available | CNA |
| Radiometer Medical Aps | AQT90 FLEX Analyzers | Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems | Application software versions >= 8.13 MR2 with Windows 10 operating system | Patch available | CNA |
- ABL90 FLEX and ABL90 FLEX PLUS Analyzers VendorRadiometer Medical Aps Affected Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems Fixed in Application software versions >= 3.5MR11 with Windows 10 operating system Patch available CNA
- ABL800 BASIC and ABL800 FLEX Analyzers VendorRadiometer Medical Aps Affected Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems Fixed in Application software versions >= 6.20MR2 with Windows 10 operating system Patch available CNA
- AQT90 FLEX Analyzers VendorRadiometer Medical Aps Affected Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems Fixed in Application software versions >= 8.13 MR2 with Windows 10 operating system Patch available CNA
Remediation
- Fix · CNA ABL90 FLEX and ABL90 FLEX PLUS Analyzers (Radiometer Medical Aps) Fixed in Application software versions >= 3.5MR11 with Windows 10 operating system
- Fix · CNA ABL800 BASIC and ABL800 FLEX Analyzers (Radiometer Medical Aps) Fixed in Application software versions >= 6.20MR2 with Windows 10 operating system
- Fix · CNA AQT90 FLEX Analyzers (Radiometer Medical Aps) Fixed in Application software versions >= 8.13 MR2 with Windows 10 operating system
Timeline
- 2026-10-02 Threatist record created from CVE-2025-14097
- 2026-10-03 Alias added: GHSA-rjfq-mp68-v4f7
References
More Radiometer Medical Aps vulnerabilities
- TI-2025-12881 Credential Disclosure vulnerability in Radiometer Products CVSS 3.1HIGH 8.4
- TI-2025-12880 Privilege boundary violation in Radiometer Products CVSS 3.1MEDIUM 6.8
Sources
- CVE-2025-14097 · CVE Program
- CVE-2025-14097 · NVD
- FIRST EPSS
- GHSA-rjfq-mp68-v4f7 · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences