Read the docs
TI-2025-12882 Aliases CVE-2025-14097GHSA-rjfq-mp68-v4f7

Remote Code Execution Vulnerability in Radiometer Products

Priority LOW 34CWE-287Published · Updated

Key facts

Attack vector
Network
Privileges required
High
User interaction
None
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
Fixed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-287.

Summary

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00465 40th percentileScored 2026-11-07
Fix Fixed ABL90 FLEX and ABL90 FLEX PLUS Analyzers (Radiometer Medical Aps): Application software versions >= 3.5MR11 with Windows 10 operating systemABL800 BASIC and ABL800 FLEX Analyzers (Radiometer Medical Aps): Application software versions >= 6.20MR2 with Windows 10 operating system+1 more product

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
No
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Radiometer Medical ApsABL90 FLEX and ABL90 FLEX PLUS AnalyzersApplication software versions < 3.5MR11 with Windows 7, Windows XP operating systemsApplication software versions >= 3.5MR11 with Windows 10 operating systemPatch availableCNA
Radiometer Medical ApsABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20MR2 with Windows 7, Windows XP operating systemsApplication software versions >= 6.20MR2 with Windows 10 operating systemPatch availableCNA
Radiometer Medical ApsAQT90 FLEX AnalyzersApplication software versions <= 8.13 MR2 with Windows 7, Windows XP operating systemsApplication software versions >= 8.13 MR2 with Windows 10 operating systemPatch availableCNA
  • ABL90 FLEX and ABL90 FLEX PLUS Analyzers VendorRadiometer Medical Aps Affected Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems Fixed in Application software versions >= 3.5MR11 with Windows 10 operating system Patch available CNA
  • ABL800 BASIC and ABL800 FLEX Analyzers VendorRadiometer Medical Aps Affected Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems Fixed in Application software versions >= 6.20MR2 with Windows 10 operating system Patch available CNA
  • AQT90 FLEX Analyzers VendorRadiometer Medical Aps Affected Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems Fixed in Application software versions >= 8.13 MR2 with Windows 10 operating system Patch available CNA

Remediation

  • Fix · CNA ABL90 FLEX and ABL90 FLEX PLUS Analyzers (Radiometer Medical Aps) Fixed in Application software versions >= 3.5MR11 with Windows 10 operating system
  • Fix · CNA ABL800 BASIC and ABL800 FLEX Analyzers (Radiometer Medical Aps) Fixed in Application software versions >= 6.20MR2 with Windows 10 operating system
  • Fix · CNA AQT90 FLEX Analyzers (Radiometer Medical Aps) Fixed in Application software versions >= 8.13 MR2 with Windows 10 operating system

Timeline

  1. 2026-10-02 Threatist record created from CVE-2025-14097
  2. 2026-10-03 Alias added: GHSA-rjfq-mp68-v4f7

References

  • TI-2025-12881 Credential Disclosure vulnerability in Radiometer Products CVSS 3.1HIGH 8.4
  • TI-2025-12880 Privilege boundary violation in Radiometer Products CVSS 3.1MEDIUM 6.8