Read the docs
TI-2025-33819 Aliases CVE-2025-53963GHSA-mmhc-m89q-9v2j

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices.

Priority MEDIUM 49CWE-521Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CISA ADP. Fix status from CNA. Impact type from CWE-521.

Summary

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00494 42nd percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
Yes
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
——all versions—No fix listedCNA
  • — Vendor— Affected all versions Fixed in — No fix listed CNA

CPE configurations NVD · 2

CPEVersionsRoleVendor
cpe:2.3:o:thermofisher:ion_torrent_onetouch_2_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableThermo Fisher
cpe:2.3:h:thermofisher:ion_torrent_onetouch_2:-:*:*:*:*:*:*:*not specified by NVDRunning onThermo Fisher
  • cpe:2.3:o:thermofisher:ion_torrent_onetouch_2_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Thermo Fisher
  • cpe:2.3:h:thermofisher:ion_torrent_onetouch_2:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Thermo Fisher

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2025-53963
  2. 2026-10-03 Alias added: GHSA-mmhc-m89q-9v2j

References

  • TI-2025-34132 An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. CVSS 3.1CRITICAL 9.8

Sources