Read the docs
TI-2025-34132 Aliases CVE-2025-54304GHSA-9r76-4gr2-pv84

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices.

Priority MEDIUM 50CWE-200Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Information disclosure
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CISA ADP. Fix status from CNA. Impact type from CWE-200.

Summary

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00535 45th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
Yes
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
——all versions—No fix listedCNA
  • — Vendor— Affected all versions Fixed in — No fix listed CNA

CPE configurations NVD · 2

CPEVersionsRoleVendor
cpe:2.3:o:thermofisher:ion_torrent_onetouch_2_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableThermo Fisher
cpe:2.3:h:thermofisher:ion_torrent_onetouch_2:-:*:*:*:*:*:*:*not specified by NVDRunning onThermo Fisher
  • cpe:2.3:o:thermofisher:ion_torrent_onetouch_2_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Thermo Fisher
  • cpe:2.3:h:thermofisher:ion_torrent_onetouch_2:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Thermo Fisher

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2025-54304
  2. 2026-10-03 Alias added: GHSA-9r76-4gr2-pv84

References

  • TI-2025-33819 An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. CVSS 3.1CRITICAL 9.8

Sources