Read the docs

Vulnerabilities

Filters

Reset all
Threatist priority
Signals
Fix status
Impact type
Attack vector
Requirements
Product

Products as named in CPE, with their records.

Published
Weakness (CWE)
2 selected

3 records

Priority LOW 20

In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an information disclosure vulnerability in the ventilator allows attackers with physical access to the configuration interface's logs to get valid checksums…

CVSS 3.1MEDIUM 4.3 Physical Low privileges No interaction EPSS 0.00293 Published 2021-03-15 No fix listed CWE-200
Priority LOW 20

In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation vulnerability in the ventilator allows privileged attackers with physical access to render the device persistently unusable by uploading s…

CVSS 3.1MEDIUM 4.3 Physical Low privileges No interaction EPSS 0.00262 Published 2021-03-15 No fix listed CWE-112
Priority LOW 23

In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the device’s configuration interface.

CVSS 3.1MEDIUM 5.2 Physical No auth No interaction EPSS 0.00282 Published 2021-03-15 No fix listed CWE-798