Read the docs
TI-2014-04431 Aliases CVE-2014-5432GHSA-7gh3-72p7-5h77

Improper authentication in Baxter SIGMA Spectrum Infusion System

Priority MEDIUM 47CWE-592Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.0 · NVD. Fix status from CNA. Impact type from CWE-287.

Summary

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Why it matters now

EPSS 0.02573 85th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
BaxterSIGMA Spectrum Infusion System6.05 (model 35700BAX) with wireless battery module (WBM) version 16.—No fix listedCNA
  • SIGMA Spectrum Infusion System VendorBaxter Affected 6.05 (model 35700BAX) with wireless battery module (WBM) version 16. Fixed in — No fix listed CNA

CPE configurations NVD · 3

CPEVersionsRoleVendor
cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*6.05VulnerableBaxter
cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*not specified by NVDRunning onBaxter
cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*16Running onBaxter
  • cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:* Versions 6.05 Vulnerable ·Baxter
  • cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Baxter
  • cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:* Versions 16 Running on ·Baxter

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2014-5432
  2. 2026-10-03 Alias added: GHSA-7gh3-72p7-5h77

References

  • TI-2020-09702 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for… CVSS 3.1CRITICAL 9.8
  • TI-2014-04432 Cleartext storage of sensitive information in Baxter SIGMA Spectrum Infusion System CVSS 3.0CRITICAL 9.8
  • TI-2020-09705 Hard-coded credentials in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
  • TI-2020-09703 Hard-coded credentials in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
  • TI-2014-04433 Hard-coded credentials in Baxter SIGMA Spectrum Infusion System CVSS 3.0CRITICAL 9.8
  • TI-2020-09699 Cleartext transmission of sensitive information in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
See all 9 →

Sources