Improper authentication in Baxter SIGMA Spectrum Infusion System
Key facts
- Attack vector
- Network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Impact type
- Security bypass
- Confidentiality · integrity · availability
- High · High · High
- Fix status
- No fix listed
Attack profile from CVSS 3.0 · NVD. Fix status from CNA. Impact type from CWE-287.
Summary
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Why it matters now
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Baxter | SIGMA Spectrum Infusion System | 6.05 (model 35700BAX) with wireless battery module (WBM) version 16. | — | No fix listed | CNA |
- SIGMA Spectrum Infusion System VendorBaxter Affected 6.05 (model 35700BAX) with wireless battery module (WBM) version 16. Fixed in — No fix listed CNA
CPE configurations NVD · 3
| CPE | Versions | Role | Vendor |
|---|---|---|---|
| cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:* | 6.05 | Vulnerable | Baxter |
| cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Baxter |
| cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:* | 16 | Running on | Baxter |
- cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:* Versions 6.05 Vulnerable ·Baxter
- cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Baxter
- cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:* Versions 16 Running on ·Baxter
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2014-5432
- 2026-10-03 Alias added: GHSA-7gh3-72p7-5h77
References
More Baxter Sigma Spectrum Infusion System Firmware vulnerabilities
- TI-2020-09702 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for… CVSS 3.1CRITICAL 9.8
- TI-2014-04432 Cleartext storage of sensitive information in Baxter SIGMA Spectrum Infusion System CVSS 3.0CRITICAL 9.8
- TI-2020-09705 Hard-coded credentials in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
- TI-2020-09703 Hard-coded credentials in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
- TI-2014-04433 Hard-coded credentials in Baxter SIGMA Spectrum Infusion System CVSS 3.0CRITICAL 9.8
- TI-2020-09699 Cleartext transmission of sensitive information in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
Sources
- CVE-2014-5432 · CVE Program
- CVE-2014-5432 · NVD
- FIRST EPSS
- GHSA-7gh3-72p7-5h77 · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences