Read the docs
TI-2014-04432 Aliases CVE-2014-5433GHSA-q82h-476f-85p3

Cleartext storage of sensitive information in Baxter SIGMA Spectrum Infusion System

Priority MEDIUM 46CWE-312Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Information disclosure
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.0 · NVD. Fix status from CNA. Impact type from CWE-312.

Summary

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Why it matters now

EPSS 0.02062 81st percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

Not available: CISA has not published SSVC decision points for this record.

Affected products

VendorProductAffected versionsFixed inStatusSource
BaxterSIGMA Spectrum Infusion System6.05 (model 35700BAX) with wireless battery module (WBM) version 16—No fix listedCNA
  • SIGMA Spectrum Infusion System VendorBaxter Affected 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 Fixed in — No fix listed CNA

CPE configurations NVD · 3

CPEVersionsRoleVendor
cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*6.05VulnerableBaxter
cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*not specified by NVDRunning onBaxter
cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*16Running onBaxter
  • cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:* Versions 6.05 Vulnerable ·Baxter
  • cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Baxter
  • cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:* Versions 16 Running on ·Baxter

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2014-5433
  2. 2026-10-03 Alias added: GHSA-q82h-476f-85p3

References

  • TI-2014-04431 Improper authentication in Baxter SIGMA Spectrum Infusion System CVSS 3.0CRITICAL 9.8
  • TI-2020-09702 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for… CVSS 3.1CRITICAL 9.8
  • TI-2020-09705 Hard-coded credentials in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
  • TI-2020-09703 Hard-coded credentials in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
  • TI-2014-04433 Hard-coded credentials in Baxter SIGMA Spectrum Infusion System CVSS 3.0CRITICAL 9.8
  • TI-2020-09699 Cleartext transmission of sensitive information in Baxter Sigma Spectrum Infusion Pumps CVSS 3.1CRITICAL 9.8
See all 9 →

Sources