Improper input validation in ICS-CERT Dräger Infinity Delta
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Impact type
- —
- Confidentiality · integrity · availability
- None · None · High
- Fix status
- No fix listed
Attack profile from CVSS 3.0 · NVD. Fix status from CNA.
Summary
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
Why it matters now
SSVC decision points
Not available: CISA has not published SSVC decision points for this record.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| ICS-CERT | Dräger Infinity Delta | Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. | — | No fix listed | CNA |
- Dräger Infinity Delta VendorICS-CERT Affected Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Fixed in — No fix listed CNA
CPE configurations NVD · 8
| CPE | Versions | Role | Vendor |
|---|---|---|---|
| cpe:2.3:o:draeger:kappa_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Dräger |
| cpe:2.3:h:draeger:kappa:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Dräger |
| cpe:2.3:o:draeger:infinity_explorer_c700_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Dräger |
| cpe:2.3:h:draeger:infinity_explorer_c700:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Dräger |
| cpe:2.3:o:draeger:delta_xl_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Dräger |
| cpe:2.3:h:draeger:delta_xl:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Dräger |
| cpe:2.3:o:draeger:infinity_delta_firmware:*:*:*:*:*:*:*:* | all versions | Vulnerable | Dräger |
| cpe:2.3:h:draeger:infinity_delta:-:*:*:*:*:*:*:* | not specified by NVD | Running on | Dräger |
- cpe:2.3:o:draeger:kappa_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Dräger
- cpe:2.3:h:draeger:kappa:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Dräger
- cpe:2.3:o:draeger:infinity_explorer_c700_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Dräger
- cpe:2.3:h:draeger:infinity_explorer_c700:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Dräger
- cpe:2.3:o:draeger:delta_xl_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Dräger
- cpe:2.3:h:draeger:delta_xl:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Dräger
- cpe:2.3:o:draeger:infinity_delta_firmware:*:*:*:*:*:*:*:* Versions all versions Vulnerable ·Dräger
- cpe:2.3:h:draeger:infinity_delta:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Dräger
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2018-19010
- 2026-10-03 Alias added: GHSA-6c7p-wgw2-hw4j
References
More Dräger Delta Xl Firmware vulnerabilities
- TI-2018-13476 Improper privilege management in ICS-CERT Dräger Infinity Delta CVSS 3.0HIGH 7.8
- TI-2019-15767 Dräger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet CVSS 4.0HIGH 7.1
- TI-2018-13478 Sensitive information in log files in ICS-CERT Dräger Infinity Delta CVSS 3.0MEDIUM 6.5
- TI-2019-15768 Dräger Infinity Delta/Kappa Patient Monitors Unauthenticated Log File Disclosure CVSS 4.0MEDIUM 5.3
Sources
- CVE-2018-19010 · CVE Program
- CVE-2018-19010 · NVD
- FIRST EPSS
- GHSA-6c7p-wgw2-hw4j · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences