Dräger SC Monitoring Devices DoS via Malformed Network Packet
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Attack requirements
- None
- Impact type
- —
- Confidentiality · integrity · availability
- None · None · High
- Fix status
- No fix listed
Attack profile from CVSS 4.0 · CNA. Fix status from CNA.
Summary
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
Why it matters now
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Partial
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Dräger | SC 6002XL | SC 6002XL | — | No fix listed | CNA |
| Dräger | SC6802XL | SC6802XL | — | No fix listed | CNA |
| Dräger | SC 7000 | SC 7000 | — | No fix listed | CNA |
| Dräger | SC8000 | SC8000 | — | No fix listed | CNA |
| Dräger | SC90000 XL | SC90000 XL | — | No fix listed | CNA |
- SC 6002XL VendorDräger Affected SC 6002XL Fixed in — No fix listed CNA
- SC6802XL VendorDräger Affected SC6802XL Fixed in — No fix listed CNA
- SC 7000 VendorDräger Affected SC 7000 Fixed in — No fix listed CNA
- SC8000 VendorDräger Affected SC8000 Fixed in — No fix listed CNA
- SC90000 XL VendorDräger Affected SC90000 XL Fixed in — No fix listed CNA
Remediation
No fixed version listed by the sources yet.
Fixed in the same release
Records fixed by the same vendor advisory or release, as the CNA lists them.
- Advisory ·
Dräger SC 6002XL, SC 7000 and 3 more products — static.draeger.com/security/download/2019-11-27-Draeger-SC7…
2 records Highest priority Priority LOW 2 LOW
Other records, highest priority first
- TI-2019-15773 Dräger SC Monitoring Devices Hard-coded Credentials and DoSPriority LOW 31
Timeline
- 2026-10-02 Threatist record created from CVE-2019-25720
- 2026-10-03 Alias added: GHSA-4wcc-pjv3-vvw7
References
More Dräger vulnerabilities
- TI-2021-09347 Draeger X-Dock Firmware before 03.00.13 has Hard-Coded Credentials, leading to remote… CVSS 3.1HIGH 8.8
- TI-2024-13029 Dräger Core 1.0.5 Denial of Service via Malformed SDC Message CVSS 4.0HIGH 8.7
- TI-2022-04579 Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering CVSS 4.0HIGH 8.8
- TI-2019-15770 Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling CVSS 4.0HIGH 8.8
- TI-2021-09348 Draeger X-Dock Firmware before 03.00.13 has Active Debug Code on a debug port, leading to… CVSS 3.1HIGH 8.8
- TI-2019-15769 Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass CVSS 4.0HIGH 8.6
Sources
- CVE-2019-25720 · CVE Program
- CVE-2019-25720 · NVD
- FIRST EPSS
- GHSA-4wcc-pjv3-vvw7 · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences