Dräger Infinity M300 VG2.x Network-Based Denial of Service
Key facts
- Attack vector
- Adjacent network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- Low
- Attack requirements
- None
- Impact type
- Denial of service
- Confidentiality · integrity · availability
- None · None · High
- Fix status
- Fixed
Attack profile from CVSS 4.0 · CNA. Fix status from CNA. Impact type from CWE-400.
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
Why it matters now
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Partial
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
Remediation
- Fix · CNA Infinity M300 (Dräger) Fixed in VG3.0
Timeline
- 2026-10-02 Threatist record created from CVE-2019-25724
- 2026-10-03 Alias added: GHSA-9492-c6v8-vj2h
References
More Dräger vulnerabilities
- TI-2021-09347 Draeger X-Dock Firmware before 03.00.13 has Hard-Coded Credentials, leading to remote… CVSS 3.1HIGH 8.8
- TI-2024-13029 Dräger Core 1.0.5 Denial of Service via Malformed SDC Message CVSS 4.0HIGH 8.7
- TI-2022-04579 Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering CVSS 4.0HIGH 8.8
- TI-2019-15770 Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling CVSS 4.0HIGH 8.8
- TI-2021-09348 Draeger X-Dock Firmware before 03.00.13 has Active Debug Code on a debug port, leading to… CVSS 3.1HIGH 8.8
- TI-2019-15769 Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass CVSS 4.0HIGH 8.6
Sources
- CVE-2019-25724 · CVE Program
- CVE-2019-25724 · NVD
- FIRST EPSS
- GHSA-9492-c6v8-vj2h · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences