Vulnerability in Baxter Welch Allyn Connex Spot Monitor
Priority MEDIUM 41CWE-1394Published · Updated
Key facts
- Attack vector
- Network
- Privileges required
- None
- User interaction
- None
- Attack complexity
- High
- Attack requirements
- None
- Impact type
- —
- Confidentiality · integrity · availability
- High · High · None
- Fix status
- No fix listed
Attack profile from CVSS 4.0 · CNA. Fix status from CNA.
Summary
Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.
Why it matters now
Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00392 33rd percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Total
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2024-1275
- 2026-10-03 Alias added: GHSA-wmcv-82jv-v932
More Baxter vulnerabilities
- TI-2024-32651 Life2000 ventilator and Service PC lack sufficient audit logging capabilities CVSS 3.1CRITICAL 10.0
- TI-2024-32650 Life2000 service tools for test and calibration do not support user authentication CVSS 3.1CRITICAL 10.0
- TI-2024-06298 Vulnerability in Baxter Connex Health Portal CVSS 3.1CRITICAL 10.0
- TI-2014-04431 Improper authentication in Baxter SIGMA Spectrum Infusion System CVSS 3.0CRITICAL 9.8
- TI-2020-09702 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for… CVSS 3.1CRITICAL 9.8
- TI-2020-09675 Hard-coded credentials in Baxter ExactaMix EM 2400 & EM 1200 CVSS 3.1CRITICAL 9.8
Sources
- CVE-2024-1275 · CVE Program
- CVE-2024-1275 · NVD
- FIRST EPSS
- GHSA-wmcv-82jv-v932 · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences