Read the docs
TI-2022-02790 Aliases CVE-2022-3027GHSA-q7x7-x9qf-5p7c

Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient…

Priority LOW 26CWE-284Published · Updated

Key facts

Attack vector
Adjacent network
Privileges required
None
User interaction
Required
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
None · None · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-284.

Summary

The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00324 25th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
No
Technical impact
Partial

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Contec HealthCMS8000 CONTEC ICU CCU Vital Signs Patient MonitorAll—No fix listedCNA
  • CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor VendorContec Health Affected All Fixed in — No fix listed CNA

CPE configurations NVD · 2

CPEVersionsRoleVendor
cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableContec Health
cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*not specified by NVDRunning onContec Health
  • cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Contec Health
  • cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Contec Health

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2022-3027
  2. 2026-10-03 Alias added: GHSA-q7x7-x9qf-5p7c

References

  • TI-2022-17808 Uncontrolled resource consumption in Contec Health CMS8000 CONTEC ICU CCU Vital Signs… CVSS 3.1HIGH 7.5
  • TI-2022-16724 Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient… CVSS 3.1MEDIUM 6.8
  • TI-2022-17783 Hard-coded credentials in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor CVSS 3.1MEDIUM 6.1
  • TI-2022-18078 Multiple binary application files on the CMS8000 device are compiled with 'not stripped'… CVSS 3.1MEDIUM 4.4

Sources