Read the docs
TI-2022-16724 Aliases CVE-2022-36385GHSA-gvx4-5vjq-6x82

Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient…

Priority LOW 32CWE-284Published · Updated

Key facts

Attack vector
Physical
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Security bypass
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-284.

Summary

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.00414 35th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
No
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Contec HealthCMS8000 CONTEC ICU CCU Vital Signs Patient MonitorAll—No fix listedCNA
  • CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor VendorContec Health Affected All Fixed in — No fix listed CNA

CPE configurations NVD · 2

CPEVersionsRoleVendor
cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableContec Health
cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*not specified by NVDRunning onContec Health
  • cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Contec Health
  • cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Contec Health

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2022-36385
  2. 2026-10-03 Alias added: GHSA-gvx4-5vjq-6x82

References

  • TI-2022-17808 Uncontrolled resource consumption in Contec Health CMS8000 CONTEC ICU CCU Vital Signs… CVSS 3.1HIGH 7.5
  • TI-2022-17783 Hard-coded credentials in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor CVSS 3.1MEDIUM 6.1
  • TI-2022-02790 Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient… CVSS 3.1MEDIUM 5.7
  • TI-2022-18078 Multiple binary application files on the CMS8000 device are compiled with 'not stripped'… CVSS 3.1MEDIUM 4.4

Sources