Read the docs
TI-2022-17808 Aliases CVE-2022-38100GHSA-w768-wqpr-555x

Uncontrolled resource consumption in Contec Health CMS8000 CONTEC ICU CCU Vital Signs…

Priority MEDIUM 41CWE-400Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
None
Attack complexity
Low
Impact type
Denial of service
Confidentiality · integrity · availability
None · None · High
Fix status
No fix listed

Attack profile from CVSS 3.1 · CNA. Fix status from CNA. Impact type from CWE-400.

Summary

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.

Why it matters now

Exploitation None reported CISA SSVC · not in CISA KEV
EPSS 0.01047 64th percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
None
Automatable
Yes
Technical impact
Partial

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Contec HealthCMS8000 CONTEC ICU CCU Vital Signs Patient MonitorAll—No fix listedCNA
  • CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor VendorContec Health Affected All Fixed in — No fix listed CNA

CPE configurations NVD · 2

CPEVersionsRoleVendor
cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*not specified by NVDVulnerableContec Health
cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*not specified by NVDRunning onContec Health
  • cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:* Versions not specified by NVD Vulnerable ·Contec Health
  • cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:* Versions not specified by NVD Running on ·Contec Health

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2022-38100
  2. 2026-10-03 Alias added: GHSA-w768-wqpr-555x

References

  • TI-2022-16724 Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient… CVSS 3.1MEDIUM 6.8
  • TI-2022-17783 Hard-coded credentials in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor CVSS 3.1MEDIUM 6.1
  • TI-2022-02790 Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient… CVSS 3.1MEDIUM 5.7
  • TI-2022-18078 Multiple binary application files on the CMS8000 device are compiled with 'not stripped'… CVSS 3.1MEDIUM 4.4

Sources