Life2000 ventilator and Service PC lack sufficient audit logging capabilities
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination.
37 records
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination.
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication.
The ventilator's microcontroller lacks memory protection.
The ventilator does not perform proper file integrity checks when adopting firmware updates.
The debug port on the ventilator's serial interface is enabled by default.
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form.
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password.
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings a…
In Baxter Connex health portal released before 8/30/2024, an improper access control vulnerability has been found that could allow an unauthenticated attacker to gain unauthorized access to Connex portal's database and/o…
In Connex health portal released before8/30/2024, SQL injection vulnerabilities were found that could have allowed an unauthenticated attacker to gain unauthorized access to Connex portal's database.
Insufficiently Protected Credentials vulnerability in Baxter Welch Allyn Configuration Tool may allow Remote Services with Stolen Credentials.This issue affects Welch Allyn Configuration Tool: versions 1.9.4.1 and prior.
Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.
The Baxter Spectrum WBM is susceptible to format string attacks via application messaging.
The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form.
The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging.
The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host.
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability.
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Ve…
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 V…
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application st…
Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access.
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Managemen…
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and n…