Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Key facts
- Attack vector
- Network
- Privileges required
- None
- User interaction
- Passive
- Attack complexity
- Low
- Attack requirements
- Present
- Impact type
- —
- Confidentiality · integrity · availability
- High · High · High
- Fix status
- No fix listed
Attack profile from CVSS 4.0 · CNA. Fix status from CNA.
Summary
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Why it matters now
SSVC decision points
CISA ADP · SSVC 2.0.3
- Exploitation
- None
- Automatable
- No
- Technical impact
- Total
The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.
Affected products
| Vendor | Product | Affected versions | Fixed in | Status | Source |
|---|---|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor | All versions | — | No fix listed | CNA |
- CMS8000 Patient Monitor VendorContec Health Affected All versions Fixed in — No fix listed CNA
Remediation
No fixed version listed by the sources yet.
Timeline
- 2026-10-02 Threatist record created from CVE-2025-0626
- 2026-10-03 Alias added: GHSA-p2j3-fm26-85wv
References
More Contec Health vulnerabilities
- TI-2025-01029 The "update" binary in the firmware of the affected product sends attempts to mount to a… CVSS 4.0HIGH 7.7
- TI-2024-11359 Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor CVSS 4.0CRITICAL 9.3
- TI-2022-17808 Uncontrolled resource consumption in Contec Health CMS8000 CONTEC ICU CCU Vital Signs… CVSS 3.1HIGH 7.5
- TI-2025-00577 Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor CVSS 4.0HIGH 8.2
- TI-2022-16724 Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient… CVSS 3.1MEDIUM 6.8
- TI-2022-17783 Hard-coded credentials in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor CVSS 3.1MEDIUM 6.1
Sources
- CVE-2025-0626 · CVE Program
- CVE-2025-0626 · NVD
- FIRST EPSS
- GHSA-p2j3-fm26-85wv · GitHub Advisory Database · CC BY 4.0
This product uses the NVD API but is not endorsed or certified by the NVD. Sources and licences