Read the docs
TI-2025-01029 Aliases CVE-2025-1204GHSA-f9x4-234p-wh9g

The "update" binary in the firmware of the affected product sends attempts to mount to a…

Priority HIGH 69CWE-912Published · Updated

Key facts

Attack vector
Network
Privileges required
None
User interaction
Passive
Attack complexity
Low
Attack requirements
Present
Impact type
—
Confidentiality · integrity · availability
High · High · High
Fix status
No fix listed

Attack profile from CVSS 4.0 · CNA. Fix status from CNA.

Summary

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

Why it matters now

Exploitation Proof of concept CISA SSVC · not in CISA KEV
EPSS 0.00483 41st percentileScored 2026-11-07
Fix No fix listed No fixed version in the sources yet

SSVC decision points

CISA ADP · SSVC 2.0.3

Exploitation
Poc
Automatable
No
Technical impact
Total

The decision (Track, Attend, Act) also depends on your mission and well-being impact; see CISA's SSVC guide.

Affected products

VendorProductAffected versionsFixed inStatusSource
Contec HealthCMS8000 Patient Monitorall versions—No fix listedCNA
  • CMS8000 Patient Monitor VendorContec Health Affected all versions Fixed in — No fix listed CNA

Remediation

No fixed version listed by the sources yet.

Timeline

  1. 2026-10-02 Threatist record created from CVE-2025-1204
  2. 2026-10-03 Alias added: GHSA-f9x4-234p-wh9g

References

  • TI-2024-11359 Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor CVSS 4.0CRITICAL 9.3
  • TI-2022-17808 Uncontrolled resource consumption in Contec Health CMS8000 CONTEC ICU CCU Vital Signs… CVSS 3.1HIGH 7.5
  • TI-2025-00526 Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor CVSS 4.0HIGH 7.7
  • TI-2025-00577 Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor CVSS 4.0HIGH 8.2
  • TI-2022-16724 Improper access control in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient… CVSS 3.1MEDIUM 6.8
  • TI-2022-17783 Hard-coded credentials in Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor CVSS 3.1MEDIUM 6.1
See all 8 →